A10 Cabinet probe error warn about malfunction of chamber probe SC.
As consequence unit stop any activity and show the message.
This message requires the intervention of an authorized technician; causes of the fault must be investigated as per prescriptions.
Description
I/O PCB detect probe signal on inlets 53 and 54 in case circuit is open or in short circuit error will occur.
Warning
Probes ar K type and generate a direct voltage so will be foundamental observe right polarity when connecting.
Reset
Error reset automatically when probe is properly connected.
Checks
Perform following checks:
- Check SC and his cabling if needed replace it
- Check probe with multimeter in ohm for closed circuit
- Check other probes is all defective you have to replace relays board.